Product Overview

Highlights

Newborn infants require a warm and cosy environment, especially so preterm babies who have an extended and urgent need of devotion, warmth and shelter. The examination and treatment areas should therefore be equipped in such a way that these small patients feel as comfortable as possible.

The baby warmer from our successful model series THERMOCARE greatly helps to achieve that aim. The compact design allows placement on any suitable table top and so easy integration into baby changing areas without fuss or major installation.

Combined with a radiant warmer of the model series CERAMOTHERM, integral workstations will provide ambient conditions equivalent to a mobile resuscitation unit.

The comfortable conductive under-pad heating warms up even hypothermic infants quickly and efficiently. The temperature is selected in increments of 0.1°C, whilst displayed and monitored electronically. Alarms will alert in the case of high or low temperature, system faults or power failure. A separate safety circuit will cut off the heating in the case of fault.


Baby warmer with gel pad, 4 rubber feet and receptacle

THERMOCARE Convenience
Order No. WY2011

Produktdetails

Dimensions T=715mm, B=600mm, H1=40mm, H2=100mm
Weight 12 kg
Operating voltage / power supply ~230 V / 50 Hz
Max. power input 160 W / 0.7 A
Self-test Yes
Temperature selection Yes, in increments of 1°C
Standard range 35°C to 37°C
Extended range 30°C to 38.5°C
Digital temperature selection display Yes
Digital actual temperature display Yes
Digital residual heat display when the device is switched-off Yes, in actual temperature display
Alarm in case of system faults Yes, visual and audible
Alarm for temperature deviation ± 1°C Yes, visual and audible
Safety cut-off at 40,5°C Yes
Power failure alarm Yes, visual and audible
Heating-up time from 24°C to 37°C Approx. 15 minutes
Colour White RAL 9002
Protection class 1
Device type B
MDD-class IIa
EN 60601-1:2006 Yes
EN 60601:1-2:2007 Yes
EN 80601-2-35:2010* Yes
  * Particular requirements for the basic safety of heating devices using blankets, pads and mattresses and intended for heating in medical use
Inspection Yearly
Preventive maintenance Yearly

Do you have questions?

Tel.: +49 (0)2207 9698-0
E-Mail: info@weyermed.com

Print this page