Product Overview

Highlights

For treatment of the neonatal hyperbilirubinemia the phototherapy device BILI-COMPACT delivers the high-effective "Blue-BAM spectrum by 10 energy saving bulbs. By the high utilization of intensity the treatment times are shortened considerably
 
The device operates in complete silence and practically no heat is emitted in the direction of the patient. Therefore it can just be placed on the canopy of warming bed THERMOCARE K, order No. WY2030 or of incubators THERMOCARE Vita order No. WY2401 and WY2402. It is also possible to place it on incubators of other brands with suitable canopy (up to 5° slant, load approx. 5 kg), without influencing the temperature control
 
For infant care systems with radiant warmer and without canopy the phototherapy device BILI-COMPACT is used with a height-adjustable stand which is mobile on four antistatic castors, two of them can be kick-stopped. The flexible swivel head allows to precisely direct the Blue-BAM radiation to the patient. So it can be brought very close to the patient and the usual radiation stray is minimized.


BILICOMPACT to be placed on warming bed and incubator

Phototherapy device with 10 energy saving lamps blue BAM/PL9/52, wave length spectrum 460 nm, integrated hour counter, rubber feet and handles.

Order No. WY1816


Mobile stand for phototherapy - when placing on a canopy is not possible

Bottom side of the device height-adjustable to 1315 to 1675 mm

Order No. WY2003

Rotation of device 360°

Vertical adjustment ± 45°

Produktdetails

General data

  • The device

    Depth 525 mm
    Width 220 mm
    Height 90 mm
    Weight 4,8 kg
    Minimum distance to the patient pad 200 mm
    Equipment of lamps (energy-saving fluorescent lamps) 10 pieces

Performance and operating data

  • Basic data

    Operating voltage / power supply ~230 V / 50 Hz
    Max. power input 140 W / 0,6 A
    Wave length spectrum 460 nm

    Bilirubin-effective radiation intensity Ebi at distance to patient pad

    350 mm 16 W/m2
    250 mm 20 W/m2
    200 mm 30 W/m2

    Irradiated area at distance to patient pad

    350 mm 400 x 700 mm
    200 mm 300 x 600 mm

Colours

Phototherapy device White RAL 9010
Base of mobile stand White RAL 9002

Technical data mobile stand

  • Mobile stand

    Depth 560 mm
    Width 560 mm
    Width 1535 to 1935 mm
    Weight 16 kg
    Necessary underpin height > 95 mm
    Castors Ø 52 mm / kickstop 4 / 2
    Bottom side of device height-adjustable 1315 to 1675 mm

Classification and standards

  • Classification

    Protection class 1
    Device type B
    MDD class IIa

    Standards

    EN 60601-1:2006 Yes
    EN 60601-2-50:2010* Yes
      * Particular requirements for the basic safety of infant phototherapy equipment

Maintenance

Inspection (recommended) Yearly
Preventive maintenance (recommended) Yearly

Do you have questions?

Tel.: +49 (0)2207 9698-0
E-Mail: info@weyermed.com


Brochure

Print this page